aV-280

Clinical Chemistry Analyzer

Compact Performance for Routine
Clinical Chemistry

Designed for efficient routine workflows, the aV-280 combines reliable analytical performance, compact dimensions and practical system architecture for modern clinical laboratories.

ϟ
200
Tests/Hour
Photometric throughput
△
150 µL Minimum reaction volume
▦
80 + 40
Positions
Reagent + sample capacity
Download Product Catalog →
Validity aV-280 Clinical Chemistry Analyzer
KEY FEATURES
⚙
Reliable Performance

Consistent results for routine clinical chemistry

↗
Compact Design

Optimized for modern laboratory environments

△
Practical Workflow

User-friendly operation and efficient processes

◇
Validated Technology

Designed for reliable and consistent operation

SYSTEM FEATURES

Designed for Efficient
Laboratory Workflows

The aV-280 integrates practical system architecture and precision components to support routine clinical chemistry testing in modern laboratories.

01

Precision Sampling System

  • External & internal mirror polish with external & internal probe washing
  • Sensitive liquid sensor with timely feedback of reagent & sample residuals
  • Collision protection and automatic probe depth adjustment
aV-280 precision sampling system
02

Optimized Mixing Performance

  • Teflon-coated mixer designed to reduce water dropping and carry-over
  • Efficient mixing performance with standardized mixing procedures
aV-280 mixing system
03

Reagent & Sample Management

  • 80 reagent positions and 40 sample positions
  • Non-stop cooling system with Peltier pad, 24-hour operation at 2°C–8°C
  • Optional barcode reader
aV-280 reagent and sample tray
04

Reaction System

  • 44 reaction cuvettes
  • Reaction volume as low as 150 µL
  • Stable and accurate reaction temperature: 37 ± 0.1°C
aV-280 reaction system
05

High-Precision Dispensing

  • Permanent ceramic piston
  • Accurate dispensing as low as 0.1 µL
aV-280 high precision ceramic syringe
06

Intuitive Operating System

  • User-friendly software interface designed for easy operation
  • Automatic sample dilution: decrease, increase and normal modes
  • Automatic checking and warning functions for reaction and calibration results
  • Linear & nonlinear calibration with multi-point calibration up to 8 points
  • Quality control support including Westgard, L-T plot and cumulative analysis
aV-280 operating software
TECHNICAL HIGHLIGHTS

Key Specifications
at a Glance

⚙
Throughput 200 Tests/Hour
△
Reaction Volume 150 µL minimum
▦
Capacity 80 Reagents
40 Samples
°C
Reaction Temperature 37 ± 0.1°C
Top